Development of a physical simulator to compare absorbed doses in medium and high-rate brachytherapy units

Authors

  • Luis Rangel Piñero Instituto Venezolano de Investigaciones Científicas, Unidad de Tecnología Nuclear, Altos de Pipe-Miranda, Venezuela
  • Carlos Rodríguez Montero Instituto Venezolano de Investigaciones Científicas, Unidad de Tecnología Nuclear, Altos de Pipe-Miranda, Venezuela

DOI:

https://doi.org/10.47187/perf.v1i36.410

Keywords:

Brachytherapy, Phantom, 137Cs, 60Co, Dosimetry

Abstract

Purpose: This is a dosimetric intercomparison study aimed at strengthening quality control in brachytherapy. The primary objective was to evaluate the agreement between the planned dose and the actual absorbed dose in clinical settings using different irradiation technologies.

Methodology: A custom-designed polymethyl methacrylate (PMMA) phantom was used for the simultaneous insertion of: a Farmer NE2571 camera, a TLD-100, EBT2 films, and PTW diodes (9192/9193). The study compared treatments using 137Cs (MDR) and 60Co (HDR) sources with a standard treatment plan of 200 cGy, analyzing the impact of calculation algorithms and the quality of diagnostic images.

Results: Significant differences were found between the planning systems and the experimental measurements. In institutions that use 137Cs, deviations of up to 51.2% were recorded, attributed to geometric uncertainty due to the obsolescence of X-ray equipment and limitations in the densitometric reading of films. The diodes enabled effective verification, detecting underestimations linked to radiation-induced sensitivity degradation.

Conclusions: The accuracy of brachytherapy depends on image quality and the use of up-to-date planning algorithms. It is imperative to standardize external audit protocols to ensure radiation safety and treatment efficacy.

Downloads

Download data is not yet available.

References

1. Urdaneta N, Vera A, Peschel RE, Wilson LD. Radioterapia oncológica: enfoque multidisciplinario. 2a ed. Caracas: Editorial Disinlimed; 2009.

2. Halperin EC, Wazer DE, Perez CA, Brady LW, editors. Perez & Brady’s principles and practice of radiation oncology. 7th ed. Philadelphia: Wolters Kluwer; 2019.

3. International Atomic Energy Agency. Calibration of photon and beta ray sources used in brachytherapy. Vienna: International Atomic Energy Agency; 2002. IAEA-TECDOC-1274.

4. International Atomic Energy Agency. Aspectos físicos de la garantía de calidad en radioterapia: protocolo de control de calidad. Vienna: International Atomic Energy Agency; 2000. IAEA-TECDOC-1151.

5. Ubrich F, Wulff J, Engenhart-Cabillic R, Zink K. Correction factors for source strength determination in HDR brachytherapy using the in-phantom method. Z Med Phys. 2014;24(2):138-52. doi:10.1016/j.zemedi.2013.08.001.

6. DeWerd LA, Kissick M, editors. The phantoms of medical and health physics: devices for research and development. New York: Springer; 2014.

7. Organismo Internacional de Energía Atómica. Determinación de la dosis absorbida en radioterapia externa: un código de práctica internacional para la dosimetría basada en patrones de dosis absorbida en agua. Rev. 1. Viena: Organismo Internacional de Energía Atómica; 2025. Colección de Informes Técnicos No. 398.

8. Knoll GF. Radiation detection and measurement. 4th ed. Hoboken (NJ): John Wiley & Sons; 2010.

9. DeWerd LA, Ibbott GS, Meigooni AS, Mitch MG, Rivard MJ, Stump KE, et al. A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: report of AAPM Task Group No. 138 and GEC-ESTRO. Med Phys. 2011;38(2):782-801. doi:10.1118/1.3533720.

10. Meertens H. In-phantom calibration of Selectron-LDR sources. Radiother Oncol. 1990;17(4):369-78. doi:10.1016/0167-8140(90)90010-T.

11. Azhari H, Hensley F, Schütte W, Zakaria G. Dosimetric verification of source strength for HDR afterloading units with Ir-192 and Co-60 photon sources: comparison of three different international protocols. J Med Phys. 2012;37(4):183-92. doi:10.4103/0971-6203.103603.

12. Thermo Electron Corporation. Model 4500 TLD reader with WinREMS: operator’s manual. Ohio: Thermo Electron Corporation; 2002.

13. Arjomandy B, Tailor R, Zhao L, Devic S. EBT2 film as a depth-dose measurement tool for radiotherapy beams over a wide range of energies and modalities. Med Phys. 2012;39(2):912-21. doi:10.1118/1.3678989.

14. León-Marroquín EY, García-Garduño OA, Camacho-López MA. Caracterización de la película radiocrómica EBT2 con diferentes sistemas ópticos. Arch Neurocien (Mex). 2013;18 Supl 1:38-43.

15. Niroomand-Rad A, Blackwell CR, Coursey BM, Gall KP, Galvin JM, McLaughlin WL, et al. Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55. Med Phys. 1998;25(11):2093-115. doi:10.1118/1.598407.

16. Fluke Biomedical. Nuclear Associates 07-444 Little Genius scanning densitometer: operators manual. Rev. 2. Cleveland (OH): Fluke Biomedical; 2005. Manual No. 07-444-1.

17. Kutcher GJ, Coia L, Gillin M, Hanson WF, Leibel S, Morton RJ, et al. Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40. Med Phys. 1994;21(4):581-618. doi:10.1118/1.597316.

18. Waldhäusl C, Wambersie A, Pötter R, Georg D. In-vivo dosimetry for gynaecological brachytherapy: physical and clinical considerations. Radiother Oncol. 2005;77(3):310-7. doi:10.1016/j.radonc.2005.09.004.

19. Allahverdi M, Jaberi R, Aghili M, Ghahremani F, Geraily G. In vivo dosimetry with semiconductors in medium dose rate brachytherapy for cervical cancer. Jpn J Radiol. 2013;31(3):160-5. doi:10.1007/s11604-012-0160-x.

20. Ghahramani F, Allahverdi M, Jaberi R. Dependency of semiconductor dosimeter responses, used in MDR/LDR brachytherapy, on factors which are important in clinical conditions. Rep Pract Oncol Radiother. 2008;13(1):29-33. doi:10.1016/S1507-1367(10)60079-X.

21. International Commission on Radiation Units and Measurements. Dose and volume specification for reporting intracavitary therapy in gynecology. Bethesda (MD): International Commission on Radiation Units and Measurements; 1985. ICRU Report 38.

22. de Almeida CE, Rodriguez M, Vianello E, Ferreira IH, Sibata C. An anthropomorphic phantom for quality assurance and training in gynaecological brachytherapy. Radiother Oncol. 2002;63(1):75-81. doi:10.1016/S0167-8140(02)00065-8.

23. Jaberi R, Babaloui S, Siavashpour Z, Moshtaghi M, Shirazi A, Joya M, et al. 3D in vivo dosimetry of HDR gynecological brachytherapy using micro silica bead TLDs. J Appl Clin Med Phys. 2022;23(9):e13729. doi:10.1002/acm2.13729.

24. Mlawa OM, Ngaile JE, Chaurasia PP, Amasi AI. Assessment of rectal dose with thermoluminescent in vivo dosimetry in high-dose-rate cobalt-60 intracavitary brachytherapy for cervical cancer: a two-arm cohort study using orthogonal images planning. J Contemp Brachytherapy. 2025;17(2):133-41. doi:10.5114/jcb.2025.150247.

25. Babaloui S, Jafari S, Palmer AL, Polak W, Sheidaei A, Shirazi A, et al. Using micro silica bead TLDs in high dose rate brachytherapy dosimetry: a phantom study. Radiat Phys Chem. 2022;191:109826. doi:10.1016/j.radphyschem.2021.109826.

26. Moura ES, Micka JA, Hammer CG, Culberson WS, DeWerd LA, Rostelato MECM, et al. Development of a phantom to validate high-dose-rate brachytherapy treatment planning systems with heterogeneous algorithms. Med Phys. 2015;42(4):1566-74. doi:10.1118/1.4914390.

27. Uniyal SC, Sharma SD, Naithani UC. Dosimetric verification of a high dose rate brachytherapy treatment planning system in homogeneous and heterogeneous media. Phys Med. 2013;29(2):171-7. doi:10.1016/j.ejmp.2012.01.004.

28. Johan S, Lobo D, Athiyamaan MS, Srinivas C, Banerjee S, Abhishek K, et al. In-vivo comparison of planned and measured rectal doses during cobalt-60 HDR CT-based intracavitary brachytherapy applications of cervical cancer using the PTW 9112 semiconductor probe. Asian Pac J Cancer Prev. 2023;24(3):897-907. doi:10.31557/APJCP.2023.24.3.897.

29. Jamalludin Z, Malik RA, Ung NM. Correlation analysis of CT-based rectal planning dosimetric parameters with in vivo dosimetry of MOSkin and PTW 9112 detectors in Co-60 source HDR intracavitary cervix brachytherapy. Phys Eng Sci Med. 2021;44(3):773-83. doi:10.1007/s13246-021-01026-x.

Downloads

Published

2026-11-23

How to Cite

Development of a physical simulator to compare absorbed doses in medium and high-rate brachytherapy units. (2026). Perfiles, 1(36), 38-52. https://doi.org/10.47187/perf.v1i36.410

Similar Articles

You may also start an advanced similarity search for this article.